An oscilloscope can be used to measure the average current over a given time interval and capture the current profile.
The plotted voltage drop can be used to calculate the current at a given point in time. The current can then be averaged or integrated to analyze current and energy consumption over a specified period.
To reduce noise, do the following:
- Use GND leads that are as short as possible.
- Use probes with 1x attenuation.
- Enable averaging mode to reduce random noise.
- Enable high-resolution function if it is available.
Use a minimum sampling rate of 200 ksps, which corresponds to one sample every 5 µs, to get a reliable average current measurement.
To accurately measure the current consumption of the nRF54LS05 DK, configure the oscilloscope to display the full range of currents to measure in the selected time window. Adjust the vertical scale in the oscilloscope to cover the full range of currents without clipping. Leave a small margin at the top or bottom of the vertical axis. Use an oscilloscope with high dynamic range of 16 bits or more to measure both small and large changes in current accurately.